All-in-one Solution for Optical Oxygen Measurement - Portavo 907 Multi Oxy
- High-resolution color graphic display
- Transflective and sunlight readable
- Li-ion battery
- Optical oxygen sensor
- Integrated pressure measurement for compensating air pressure fluctuations
- Micro USB port and Paraly SW 112 software
- Quantity:
- - +
Product Specification
Portavo 907 Multi Oxy
The all-in-one solution for optical oxygen measurement. Multiparameter device for digital pH, conductivity and oxygen sensors with Memosens technology
The only portable device for all Memosens parameters. The powerful Li-ion battery can be charged in the device via USB. The clear network diagram provides an at-a-glance view of the sensor condition.
The data logger for up to 10,000 entries r-ecords measuring point, annotation, sensor ID, sensor serial number (Memosens), primary value, temperature, time stamp and device status.
User-friendly software
Portavo 907 proves that a high level of functionality and easy use are not mutually exclusive. It proceeds step by step through the calibration procedure. Technical terms are clearly explained in the context-- sensitive help.
Portavo 907 Multi Oxy Optical Oxygen Measurement Facts
- High-resolution color graphic display
- Transflective and sunlight readable
- Li-ion battery
- Optical oxygen sensor
- Integrated pressure measurement for compensating air pressure fluctuations
- Micro USB port and Paraly SW 112 software
- A sensor quiver protects the sensor from damage and drying out
- The high-performance polymer housing ensures low water absorption and high impact resistance
- Intelligent data logger with 10,000 entries and graphical representation
- IP 66/67 protection
- The mineral glass display is perfectly readable even after years
Comprehensive data logger
The following logging types can be selected:
- Manual logging
- Time-controlled logging at a fixed interval
- Signal-controlled logging of measured value and temperature
- Combined time- and signal-controlled logging
- Threshold-controlled logging with pre-trigger
Portavo 907 Multi Oxy Optical Oxygen Measurement Specifications
Optical oxygen input
|
M12 socket for SE 340 sensor or Memosens sensors | ||
OXY meas. ranges at 20 °C
|
Saturation | 0.000 ... 200.0 % | |
Concentration | 000 μg/l … 20.00 mg/l | ||
Partial pressure | 0.0 … 1000 mbars | ||
Response time
|
t90 < 30 s | ||
t99 < 60 s | |||
Measurement error | Zero signal < 0.1 % full saturation | ||
Measuring cycle | Approx. 1 s | ||
Measurement error | < 0.1 % meas. val. | ||
Temperature meas. range | 0.0 … 50.0 °C | ||
Measurement error | Temperature ± 0.2 K | ||
Sensor standardization
|
Automatic calibration in air, humidity adjustable | ||
Zero calibration | |||
Max. overpressure | 2.5 bars | ||
Immersion depth
|
min. 60 mm | ||
max. 25 m | |||
Storage Temperature input
|
Quiver | ||
2 x 4 mm dia. for integrated or separate temperature detector | |||
Measuring ranges
|
NTC 30 kΩ | -20 … +120 °C | |
Pt 1000 | -40 … +250 °C | ||
Measuring cycle | Approx. 1 s |
|
|
Measurement error1,2,3) < 0.2 K (Tamb = 23 °C) | TC < 25 ppm/K | ||
Memosens pH input (also ISFET)
|
M8 socket, 4 pins, for Memosens lab cable and M12 socket, 8 pins | ||
Display ranges
|
pH | -2.000 … +16.000 | |
mV | -2000 … +2000 mV | ||
Temperature | -50 … +250 °C | ||
Memosens ORP input
|
M8 socket, 4 pins, for Memosens lab cable and M12 socket, 8 pins | ||
Display ranges
|
mV | -2000 … +2000 mV | |
Temperature | -50 … +250 °C | ||
Sensor standardization | ORP calibration (zero adjustment) | ||
Permissible calibration range |
ΔmV (offset) |
-700 … +700 mV | |
Sensor standardization
Operating modes
Calimatic buffer sets
Permissible calibration range
|
pH calibration | ||
Calimatic | Calibration with automatic buffer recognition | ||
Manual | Manual calibration with entry of individual buffer values | ||
Data entry | Data entry of zero and slope | ||
Knick CaliMat | Ciba (94) | User defined | |
NIST technical | HACH | Mettler-Toledo | |
NIST standard | Hamilton | WTW techn. buffers | |
DIN 19267 | Reagecon | ||
Zero point | 6 ... 8 pH | ||
With ISFET: | –750 … +750 mV operating point (asymmetry) | ||
Slope | Approx. 74 … 104 % | ||
Calibration timer*)
Sensoface
|
Interval | 1 … 99 days, can be switched off | |
Provides information on the sensor condition | |||
Evaluation of | zero/slope, response, calibration interval |
If you cannot find what you want, you can entrust ISweek to source for you. Just click:
Sourcing ServiceRecommended Products
-
Smart Aqua Meter for pH, ORP or Conductivity and Temperature Meter
-
Loop-Powered, Isolated Toroidal Conductivity Transmitter
-
Menu Driven Loop-Powered Toroidal Conductivity Transmitter
-
Loop-Powered Contacting Conductivity Transmitter
-
Loop-Powered Dissolved Oxygen Transmitter
-
pH/mV Transmitter
-
Multi-parameter (COD / BOD / TOC) Water Quality Automatic Analyzer
-
Cabinet Mounted Turbidity / TSS / MLSS Analyzers
-
Cabinet Installed Chlorine Analyzer (Polarography)
-
On-line TRO Analyzer (DPD Spectrophotometry)