Limiting Current Type Zirconia Oxygen Sensor - SO-E2-020

? TO8 Housing
? Measuring gas: Oxygen concentration
? Measuring medium: Gaseous atmosphere
? Measuring principle: Limiting current type sensor
? Measuring ranges: 0.01 - 2.0 vol.% O2
? Output current: 150 μA - 250 μA
? Accuracy: ±200 ppm O2
? Reproducibility: < 100 ppm O2

  • Model Number: SO-E2-020
  • Data Sheet: Date Sheet File
Quantity:
- +
  • Limiting Current Type Zirconia Oxygen Sensor
  • Limiting Current Type Zirconia Oxygen Sensor
Mouse over image to zoom
Share:

Product Specification

SO-E2-020 Limiting Current Type Zirconia Oxygen Sensor Operational Principle
When voltage is applied to a zirconia electrolyte cell, oxygen is pumped through the zirconia disc from the cathode side to the anode side because the carriers of the current flowing through the zirconia electrolyte are oxygen ions. By attaching a cap with a pinhole on the cathode side of the cell and by increasing the voltage over the cell the current shows saturation due to the rate limiting step in the transfer of oxygen to the cathode. This saturation current is called limiting current and is nearly proportional to the ambient oxygen concentration.

SO-E2-020 TO8 Housing, ? 15.3 mm; H= 13.7 mm; pin distance 7.18 mm


Below the advantages of the Zirconia Oxygen Sensor SO-E2-020:
? Measuring range 0.01 to 2.0 vol.% oxygen
? High accuracy
? For many types a more or less linear characteristic
? Small temperature dependence of the sensor signal
? Low interference with other gases
? Long service life
? In many cases "Single point calibration" necessary only once


SO-E2-020 Limiting Current Type Zirconia Oxygen Sensor Applications:
Medical
  ? Oxygen Concentrators
  ? Incubators
Laboratory
  ? Inert gas processing cabinets (Glove boxes)
  ? Incubators (controlled bacterial growth)
Food industry
  ? Packaging
  ? Controlled food testing
  ? Monitoring fruit ripening processes (storage / transport)
Household/Gastronomy
  ? High-temperature humidity measurements (> 100°C) for baking / roast automation
Measuring technique
  ? Oxygen Meters (stationary / portable)
  ? Measurements under controlled O2 content
  ? Air conditioning and ventilation 
Security technology/Monitoring
  ? Fire protection (increased N2 atmosphere e.g. Server rooms)
  ? Greenhouses, wine cellar
  ? Gas storage, refineries
  ? Diving
  ? Fermentation units
(Electrical-) industry
  ? Inert gas processing machines and cabinets
  ? Inert gas welding monitoring
  ? Storage with increased N2 atmosphere (oxidation prevention)
  ? Drying units
  ? Nitrogen concentrators
  ? Exhaust gas measurement


SO-E2-020 Limiting Current Type Zirconia Oxygen Sensor 
Characteristic Data:

Measuring gas
Oxygen concentration
Measuring medium
Gaseous atmosphere
Measuring principle
Limiting current type sensor
Measuring ranges
0.01 - 2.0 vol.% O2
Response time (t90)
2 to 25 sec. (Depends on: sensor type, gas flow, measuring chamber)
Sensor voltage / heating voltage / power consumption / heater cold resistance
   Sensor voltage:
   Heating voltage:
   Power consumption:
   Cold resistance:
0.7 to 1.6 Volt
3.6 to 4.4 Volts
1.3 to 1.8 Watts (depends on application and packaging)
R(25°C) = 3.25Ω ± 0.20Ω
Warm up time
Min. 30 s
Maximum permissible 
operating temperature

350°C (*)
*Depending on the cable and filter assembly (see paragraph: specification, Cable assembly)
Permissible volumetric flow 
rate (Purging the sensor)

For all sensor types, the maximum flow rate depends on the way of purging the sensor 
(sensor in direct gas flow, gas beam shape, etc.) and the size of the measuring chamber. 
Exception: SO-E1-xxx (TO8 + hose connection for direct gassing). Permissible flow 
[ml/min]: 100 - 500 (optimal: 250)

Lifetime (MTTF)
 ~ 20.000 hours (*)
*Depending on measuring medium. Stated lifetime refers to a heated, operated sensor of 
type SO-xx-250 and SO-xx-960. A failure or reaching the lifetime typically means a slight 
deviation from the dispatch specifications.

Vibration resistance
Sensors based on the TO8 socket as well as those built with a TO39 socket (SO-Bx-xxx, 
SO-Ex-xxx, SO-Ax-xxx) meet the European Norm EN60068-2-6 (Sinusoidal vibration tests).


If you cannot find what you want, you can entrust ISweek to source for you. Just click:

Sourcing Service
亚洲精品制服丝袜四区| 国产亚洲精品影视在线产品| 亚洲精品乱码久久久久久久久久久久 | 亚洲2022国产成人精品无码区| 国产馆精品推荐在线观看| 久久99精品福利久久久| 国产成人精品视频一区| 久久精品国产亚洲AV蜜臀色欲 | 人人妻人人澡人人爽人人精品浪潮 | 久久亚洲AV无码精品色午夜| 久久精品国产99国产精品| 国产精品一卡二卡三卡四卡| 99精品免费观看| 伊人精品视频一区二区三区| 精品久久久久久中文字幕无码| 久久午夜精品视频| 精品久久久久久无码人妻蜜桃| 69精品人人人人人人人人人 | 91精品久久久久久久久中文字幕 | 久久精品国产一区二区三区| 国产精品高清全国免费观看| 国产福利在线观看精品| 无码精品视频一区二区三区| 精品久久久久久久中文字幕| 国产精品不卡在线| 亚洲精品天堂无码中文字幕| 久久久精品人妻一区二区三区四| 极品精品国产超清自在线观看| 国产99精品在线观看| MM1313亚洲精品无码久久| 国产成人精品免费视频大| 午夜精品乱人伦小说区| 宅男在线国产精品无码| 国产成人精品久久| 精品日韩二区三区精品视频| 精品久久久久成人码免费动漫 | 国产精品老女人精品视| 国产精品1区2区3区在线播放| 亚洲av永久无码精品国产精品| 9久热精品免费观看视频| 四虎精品免费永久免费视频|