Digital Gas Flow Sensor - MFM2070U

? Flow Range: 0.07 ~ 70 SLM
? Operating Temperature Range: -40 ~ 85°C
? Accuracy: +/-2.5% Reading
? Resolution?: 15bit
? Sampling Time: 7.5ms
? Low Power Consumption
? Digital I2C

  • Model Number: MFM2070U
  • Data Sheet: Date Sheet File
Quantity:
- +
  • Digital Gas Flow Sensor
  • Digital Gas Flow Sensor
Mouse over image to zoom
Share:

Product Specification

MFM2070U Digital Gas Flow Sensor DESCRIPTION
MFM2000 series is a gas flow sensor with low pressure drop capable of measuring medical gases up to 70 lpm. It is configured to be customizable with different flow inlet and outlet configurations.

MFM2070U Digital Gas Flow Sensor FEATURES
? Low Pressure Drop Flow Sensor
? High Accuracy +/-2.5% Reading
? Calibrated and Temperature Compensated for Air
? Low Power Consumption
? Digital I2C
? Made for Medical Applications 
? Calibrated for Air, and O2

MFM2070U Digital Gas Flow Sensor FLOW PERFORMANCE
Measurements performed with air at 23°C, 1 Atm, 50% RH, at 3.3V DC power supply, unless otherwise specified. Calibration conditions for standard liters per minute slm: 20°C, 1013mbar.

Parameter
Condition
Min
Typical
Max
Unit
Flow Range
Calibration gas: Air
0.07
70 SLM
Calibrated Temperature Range

0
50 °C
Operating Temperature Range

-40 
10 85 °C
Storage Temperature Range

-40 

105 °C
Accuracy
25oC

+/-2.5

% Reading
Temperature Error
0-50oC


0.1 reading/°C
Resolution
I2C

15
bit
Sampling Time


7.5

ms
Power Consumption
@ Recommended Heater
Power at zero Flow


7.0 mA
PD


650 uA
Pressure Drop
Max flow without mesh


120 Pa
Max flow(with mesh screen)


400 Pa
Operating Pressure Range
Absolute
0.7
1.3 Bar
Operating Overpressure
Absolute

+/-0.2

Bar
Averaging
10 frames running window averaging
Wetted Materials
PES (Medical grade: biocompatible; ISO 10993 or USP Class VI), FR4, Silicon
Nitrite, Silicon, Epoxy, Gold.
Standards
RoHS, REACH, ISO15001, ISO10993-1, ISO10993-5,UL94 V0, IEC

ELECTRICAL SPECIFICATIONS (VDD=3.3V, TA=25C, UNLESS OTHERWISE NOTED)
Parameter
Condition
Min
Typical
Max
Unit
Supply DC Voltage
DC
2.7 3.3 5.5 V
Power Consumption
@ Recommended Heater Power at
zero Flo


7.0 mA
PD


650 uA
Data format Scaling Factor
User read count value from register 0,
then multiply by 2.5 to get actual
flow rate in sccm
2.5
sccm

MFM2070U Digital Gas Flow Sensor MECHANICAL SPECIFICATIONS
Specification
Condition
Min
Typical
Max
Unit
Burst Pressure
Over ambient
1

Bar

If you cannot find what you want, you can entrust ISweek to source for you. Just click:

Sourcing Service
亚洲精品无码av人在线观看| 久久国产精品免费一区二区三区| 91精品91久久久久久| 精品国产污污免费网站| 久久青青草原精品国产软件| 在线观看91精品国产网站| 免费看国产精品3a黄的视频| 99久久久国产精品免费无卡顿| 久久99精品久久水蜜桃| 精品四虎免费观看国产高清午夜| 香蕉在线精品视频在线观看2| 国产高清一级毛片精品| mm1313亚洲精品国产| 国产精品亚洲色图| 国产精品视频公开费视频| 少妇人妻偷人精品视蜜桃| 国产成人精品第一区二区| 中文字幕动漫精品专区| 久久久久久久久久久精品尤物| 三上悠亚久久精品| 久久永久免费人妻精品| 久久精品国产亚洲AV麻豆~| 亚洲av永久无码精品国产精品 | 国产香蕉精品视频在| 三上悠亚日韩精品| 精品国产一区二区22| 久久97久久97精品免视看| 国产偷国产偷高清精品| 精品无码人妻久久久久久| 国产精品人人做人人爽| 国产精品一区12p| 亚洲精品无码专区在线在线播放| 国产色无码精品视频免费| 亚洲色精品vr一区二区三区| 精品一卡2卡三卡4卡免费视频| 久久精品九九亚洲精品天堂| 久章草在线精品视频免费观看| 亚洲国产精品久久久久| 精品一区二区三区无码免费视频 | 国产成人亚洲综合无码精品| 亚洲av永久无码精品网站|