Flexible Sampling System - S2K

Dimensions: 345(H) x 292(W) x 127(D) mm; (13.6"H x 11.5"W x 5.0"D).
Weight: 5.5 Kg (12 Lbs).
Relay: SPST Normally Open, 5 amp resistive. 220 VAC max. 1/8 HP motor.
Internal Pump: Oscillating AC Diaphragm. Replaceable.
Flow Meter: Dwyer Instrument, Inc. RMA-13-SSV. Ranges as specified.
Power: 110/220 VAC, 50/60 Hz.
Temperature Range: -20oC to 50oC.
Humidity: Non-condensing.

  • Model Number: S2K
Quantity:
- +
  • Flexible Sampling System
Mouse over image to zoom
Share:

Product Specification

S2K Flexible Sampling System


Model S2K Sampling System offers the following flexible features:

    ? Basic continuous sampling

    ? Basic periodic sampling with user adjustable time interval

    ? Sample dilution to allow sensors to monitor very high gas concentrations

    ? Timer-controlled external bypass pump which improves sampling time, simplifies sample filtration requirements, and reduces pump maintenance

    ? Provide viable solution for difficult applications


Description:
The model S2K Sampling System is a flexible sampling system which is designed to condition the sample as required. This system not only brings the sample to the sensor, in many cases, it is the only viable solution to difficult applications. For instance, in applications where the sample concentration is too high for the sensor, the model S2K's ability to dilute samples allows the sensor to be exposed to the sample and provide an accurate reading. Another advantage of non-continuous sampling is the fact that the zero point is confirmed on every reading which increases the confidence level of the measurement. Fig. 1 illustrates phosphine gas diluted at a 5:1 ratio measured on a solid state sensor.


Functions:
The S2K controls the following functions:
1. A 3-way solenoid valve: A DIP switch is used to set the amount of time the sample is to be exposed to the sensor. The rest of the time the sensor is exposed to clean air.
2. An external pump control timer which sets the on-off time of the pump. The purpose of the external pump is to transport the sample in larger quantities which shortens the transport time. Additionally, it is easier to draw a controlled amount of sample to be conditioned before exposing it to the sensor. The timer turns the pump on only as needed which prolongs the pump life and reduces pump maintenance.
3. A smaller, internal pump is on constantly to flow either sample or clean air to the sensor.
There are three timers required to implement the complete cycle as shown in Fig. 2 below.



Fig. 2


1. External pump delay: This is the time delay from the beginning of the cycle before the external pump is activated. During this time, the external pump is off. The timer can be set from 0-255 minutes, with a resolution of 1 minute.
2. External Pump Lead Time: This is the amount of time the external pump is "ON", before the solenoid valve is switched to let the sample into the sensor chamber. The timer can be set from 0-2550 seconds, with a resolution of 10 seconds. 

3. This is the amount of time the solenoid valve is activated to allow sample flow into the sensor chamber. The timer can be set from 0-2550 seconds, with a resolution of 10 seconds.


Each timer is set by an eight position DIP switch with the following count values:

1 + 2 + 4 + 8 + 16 + 32 + 64 + 128 = 255. This means that position one of the DIP switch has a value of "1" while position 8 has a value of "128". The time which is set for the timer is calculated by taking the count value and multiplying it by the resolution of the timer. A time of zero is achieved by having all of the switches in the open position.


There is a micro-switch which activates a relay contact if there is a failure of the flow into the sensor chamber. A calibration switch with an LED indicator allows the user to perform standard calibration functions on the sensors. The S2K housing is designed to be used with standard IST 4-20 IQ sensor modules but other sensor modules (such as the SM95) can be used also.


The S2K sampling system can be configured with a variety of options. Fig. 3 illustrates the two-stage sampling system. Figure 4 shows a complete system with a sample diluting assembly. IST will assemble the system to meet application requirements.


System Specifications:
Enclosure: NEMA 4X.
Dimensions: 345(H) x 292(W) x 127(D) mm; (13.6"H x 11.5"W x 5.0"D).
Weight: 5.5 Kg (12 Lbs).
Timer controller function: See description above.
Relay: SPST Normally Open, 5 amp resistive. 220 VAC max. 1/8 HP motor.

    Three relays are provided for the following:

     1. No flow into sensor chamber.
     2. External pump activation Relay.
     3. Solenoid valve.
Internal Pump: Oscillating AC Diaphragm. Replaceable.
Flow Meter: Dwyer Instrument, Inc. RMA-13-SSV. Ranges as specified.
Power: 110/220 VAC, 50/60 Hz.
Temperature Range: -20oC to 50oC.
Humidity: Non-condensing.

Fig. 4


If you cannot find what you want, you can entrust ISweek to source for you. Just click:

Sourcing Service
国产成人一区二区三区精品久久 | 久久久久久久91精品免费观看 | 精品成在人线AV无码免费看| 国产高清在线精品一本大道国产| 精品国产专区91在线尤物| 久久国产精品久久精| 精品一区二区三区色花堂| 国产精品美女视视频专区| 真实国产乱子伦精品免费| 亚洲成人国产精品| 在线观看免费精品国产| 精品熟人妻一区二区三区四区不卡| 国产精品一区二区av| 国产精品1024视频| 国产精品一卡二卡三卡| 精品国产三级a∨在线观看| 1000部精品久久久久久久久| 99re在线视频精品| 久久久久久亚洲精品| 亚洲精品无码成人片久久| 国产精品毛片无码| 精品久久中文字幕| 网友自拍区视频精品| 久久精品国产亚洲77777| 国产精品国色综合久久| 97久久国产露脸精品国产| 国产精品国产三级在线高清观看 | 国产精品乱码一区二区三| 精品国产免费一区二区| 色妞www精品视频免费看| 一本久久a久久精品综合香蕉| 国产精品视频九九九| 人人鲁人人莫人人爱精品 | 无码精品人妻一区二区三区免费看| 久久国产精品免费看| 中国精品videossex中国高清| 精品国产一区AV天美传媒| 久久亚洲精品11p| 四虎永久在线精品国产馆V视影院| 国产精品成人一区二区| 国产精品香港三级国产AV|